Condition Desc. Used
Manufacturer HP
Volts 115
Model HP1047A
Length 17
Width 16
Height 6.5
Weight 44
Phase 1

Flat Shipping Rate: $45


This item will to ship anywhere in the contiguous US for the above rate. Call for international or combined rates.

RI Detector, RID, HPLC Refractive index detector, HP1047A, HP Agilent

Mfg: HP

Mdl: HP1047A

$1,650.00

Quantity: 1

Inventory #: 5574

Call 610-286-1655 with questions or shipping quotes.
REQUEST MORE INFO

This is a used refractive index detector made by HP, model HP1047A.  We powered up the unit as can be seen in the photos but did not actually take measurements with the instrument.  The heater works and achieves a setpoint which is indicated by a flashing green heater light.  The  light insensity error light is on which according to the manual could be solved by flushing the unit.  Light intensity is the relationship between two photodiodes.  Currently TP1 is reading  -1.32 and TP2 is -2.34 and the average is -1.83.  The gross zero adjustment buttons were pressed which activates a motor which moves a mirror inside.  The unit is in good overall condition but it does have a light intensity error which could be as simple as a system flush and calibration. 

 

  • Includes manual
  • Refractive index range: 1 to 1.75
  • Measurement: 1/64 x 10-5 delta RIU to 32 x 10-5 delta RIU
  • Max pressure: 7 bar